6.0 SPECIFICATIONS
6.1 CONTROLLER
Operating Range: BFS-3MP and 5MP: -35°C to ambient
BFS-30MP: -40°C to ambient
Digital Readout Resolution: 0.1°C
Accuracy: ±0.2°C
Ambient Operating Range: 15° - 45°C
Input Power Requirements: 100 - 120VAC @ 4 Amps or
200 - 240VAC @ 2 Amps
Grounded Outlet
Power Output to stage: BFS-30MP - 12VDC @ 10Amps
BFS-3 & 5 MP - 5VDC @ 10Amps
Display: Dual LED display of set and run Temperature
Safety Features: Power shutoff during over-temperature condition caused by
inadequate coolant flow (50°C). Short circuit protection.
Size: 9 1/2” wide x 10” deep x 4 1/2” high
Weight: 5 lbs
Other Features: Self-sealing water connectors.
Auxiliary switched AC output for ECV-1 or PTU-3.
6.2 STAGE
Temperature Range: -40°C to ambient
Controlled Surface Area: BFS-3MP - 3 x 4cm
BFS-5MP - 3.8 x 4cm
BFS-30MP - 7.5 x 8cm
Weight: 15ozs, including leads
Lead Length: 5ft
8
EB BFS-MP 6310 04/04
Vista de pagina 4
1 2 3 4 5 6 7 8 9 10 11 12

Comentarios a estos manuales

Sin comentarios